Data storage parameters properties – MTS Fracture Template User Manual

Page 50

Advertising
background image

Description

Parameter

Specify the rate at which the system reloads
the specimen to the load value at the

JIC Reload Rate

beginning of the first unload segment in each
step increment. If the ramp rate is invalid for
the specimen being tested, the application
corrects the value of the ramp rate when you
click Run to start the test.

Specify the method for determining the
amount of force to apply in both the unload
and reload segments:

JIC Unload Method

Absolute—Specify that the amount force
to apply is in absolute terms.

Percentage—Specify that the amount of
force to apply is a percentage of the
increment peak load.

Minimum of absolute and
percentage
—Specify that the amount
force to apply is the smaller resulting
value of the two methods above.

Specify the rate at which the system unloads
the specimen in the unload segments in each

JIC Unload Rate

step increment. If the ramp rate is invalid for
the specimen being tested, the application
corrects the value of the ramp rate when you
click Run to start the test.

Data Storage Parameters Properties

Data Storage Parameters

Description

Parameter

Specify the percentage of precrack activity
that must complete before data is saved to
disk.

Precrack Save Percent
Limit

Specify the cycle intervals written to disk
during the precrack activity.

Store Every Nth Precrack
Cycle

Specify the amount of crack length change
that will cause data to be stored to disk.

Crack Size Change Store

Typically, operators will not store every cycle,

50 | MTS TestSuite

JIC Fracture Toughness Template

Advertising