Kic parameters, Data storage parameters properties – MTS Fracture Template User Manual

Page 64

Advertising
background image

Description

Parameter

Specify the percentage of the measured load
range used as the upper limit for linear least

Precrack Upper Least
Squares Fit Percentage

squares fit compliance determination. This
parameter is not used for FCG DCPD tests.

Specify the percent of maximum load level
at which the voltage is measured. The load

Precrack Measure Load
Level Percent

is applied at this level to ensure the crack is
open. All voltage data below this load will not
be used to calculate crack length. This
parameter is used only for FCG DCPD tests.

Specify the number of cycles at which
precrack activity stops.

Precrack Cycle Limit

Specify the maximum stress intensity desired
at the end of precrack activity. The initial

Precrack Final Maximum
K

precrack Maximum K is 1.4 times the Final
Maximum K value.

Select Yes to cause an interlock at the end
of precrack. This can shut down hydraulics,
if desired.

Shutdown Station at End
of Precrack

KIC Parameters

In the KIC Parameters section of the main menu, you specify the KIC Force Ramp Rate, which is the rate
for the ramp segment of the KIC step.

Data Storage Parameters Properties

Data Storage Parameters

Description

Parameter

Specify the percentage of precrack activity
that must complete before data is saved to
disk.

Precrack Save Percent
Limit

Specify the cycle intervals written to disk
during the precrack activity.

Store Every Nth Precrack
Cycle

Specify the amount of crack length change
that will cause data to be stored to disk.

Crack Size Change Store

64 | MTS TestSuite

KIC Fracture Toughness Template

Advertising