4.2.0 Install a Specimen

4.2.1.0 Before You Begin

Before you can install a specimen, you must open a test station and apply hydraulics. For more information,
see

Starting Station Manager and Loading Station Configuration

on page 18 and

Starting MPX and Applying

Hydraulic Power

on page 19.

4.2.2.0 Setting Station Limits in Station Manager Application

To reduce the risk of injury when installing specimens, set station limits that limit the force applied by the
actuator. After installing the specimen, you may have to adjust these limits to allow the test to run.

Note:

For more information about setting limits, see MTS Series 793 Control Software.

1. Maximize the Station Manager window.

Warning:

Actuators can produce dangerous forces.

A moving actuator can injure anyone in its path.

Always set station force limits before installing a specimen.

2. Set the station limits and the resulting action.