Facility Preparation

Site Preparation

Relative Humidity

The recommended relative humidity for the test room is within the range of 5%
to 85% (non-condensing). The risk of static discharge, which easily damages
logic components and causes loss of data in memory devices, is increased by low
humidity. Excessive humidity can result in electrical leakage currents or
component failure.

Leveling System Components

The load frame is mounted to the foundation, so no leveling component is
needed. The Integrated Operations Platform has leveling feet that allow the
cabinet to be leveled as necessary. No particular level alignment of the Integrated
Operations Platform is necessary other than to ensure the footing is solid and the
Integrated Operations Platform cannot be rocked.