Integrated operations platform specifications, Integrated operations platform specifications 16 – MTS Criterion Series 60 User Manual

Page 16

Advertising
background image

MTS Criterion™ Series 60

16

System Specifications

Site Preparation

Integrated Operations Platform Specifications

The following tables provide the specifications for the Integrated Operations
Platform.

General specifications

Model-specific

specifications

P

ARAMETER

S

PECIFICATION

Environmental

Operating temperature

Humidity

Noise

rating at 1 m

For use in a controlled environment

5–40°C (41–104°F)

5–85% noncondensing

70 dB (A) fully compensated

Hydraulic fluid

Reservoir capacity

Exxon Mobil DTE 25

85 L (22.5 gal) maximum

Electrical power

Line voltage

Control voltage

3~3W+PE, 400 V 50 Hz or 480 V 60 Hz

24 V DC

Dimensions

Height

Length

Width

1900 mm (75 in)

1040 mm (41 in)

720 mm (28 in)

P

ARAMETER

64.305 /64.305EL S

PECIFICATION

64.605/64.106/64.605EL/64.106EL

S

PECIFICATION

Pump/motor

Line voltage starter configuration

Line voltage starter configuration

Maximum continuous
pressure

23 MPa (3300 psi)

23 MPa (3300 psi)

Maximum flow capacity

3.3 L/m (0.9 gpm) at 50 Hz

4.0 L/m (1.0 gpm) at 60 Hz

5.1 L/m (1.3 gpm) at 50 Hz

6.2 L/m (1.6 gpm) at 60 Hz

Motor rating

1.5 kW (1.6hp) at 50/60 Hz

2.2 kW (2.9 hp) at 50/60 Hz

Heat load (maximum)

12.3 kW (42,000 Btu/hr)

20.5 kW (70,000 Btu/hr)

Weight with maximum oil
85 L (22.1 gal)

480 kg (1058 lb)

480 kg (1058 lb)

Advertising