Software – Retsch PM GrindControl User Manual

Page 20

Advertising
background image

Software

20

After 15 minutes without assignment to a measurement task, the wireless
grinding jar cover switches off. If the status indicator lamp is not glowing red
any longer, take out one of the batteries and then put it back in again.

2.

Click on the Add measurement system

button.

Once the system that corresponds to the MAC address has been found,

3.

confirm with the OK button.

The status indicator lamp now glows green.

The “System Information” program window is displayed.

4.

In the Name field, enter a name for the wireless grinding jar cover.

Fig. 15: Name field

5.

After clicking on the Accept button,

6.

you can end the procedure by clicking on the Close button.

The changed name is the name that is stored in the measurement system. On
connecting the next time, the measurement system will identify itself with this new
name.

7.

If one or more measuring systems are logged in with the MAC address
already, these are displayed in the “Configure Measuring Systems” program
window:

Inactive Systems:

The measuring system is switched off or is outside the transmitter range.

Active Systems:

The measuring system can be added to the measurement task.

If you switch on a measuring system at a later point in time or bring it into the
transmitter range, this is not automatically displayed in the list of active systems.
With the Search for Known Measuring Systems button you can start a new
search for measuring systems.

Fig. 10: Search for Known Measuring Systems button

8.

Open the active system by double-clicking on it and then click on the Assign
field.