Software
21
Fig. 11: Assign the active system to the measurement
This adds the wireless grinding jar cover as one of the four possible measurement
systems for the current measurement task.
9. After clicking on the Accept button, you can end the procedure by clicking on
the Close button.
10. Now click on the Continue button in the Configure Measuring Systems
program window.
In the “Measurement Task” window, you can input a remark for the measurement
task.
11. Click on the Continue button.
12. in the “Measurement File” a name must be inputted for the first
measurement of the measurement task.
You can also input the operator and a remark for the measurement in the
“Measurement File” window.
13. You can end the creation of the measurement task by clicking on the Finish
button.
5.5.1.2 Menu File - Open Measurement Task
Use this command to open an existing measurement task. Another
measurement will be added.
Alternatively, you can also use the following button.
Fig. 12: Add New Measurement File to a Measurement Task button
Once you click on this symbol or the command, the “Open” window opens.
1.
Select the measurement task to which you wish to add another
measurement and click on Open.
The "Search for Assigned Measuring Systems" opens and you can see the
list of the inactive systems.
If you switch on a measuring system at a later point in time or bring it into the
transmitter range, this is not automatically displayed in the list of active systems.
You can use the Search for Known Measuring Systems button to start a new
search for measuring systems.
Fig. 13: “Search for Known Measuring Systems” button
If you double-click on the active measuring systems, information on these systems
will be displayed.
2.
Click on the Continue button.
The measurement task window opens and displays which measuring
systems are involved in the new measurement.
3.
Click on the Continue button.
The Measurement File window opens.