Wireless grinding jar cover – Retsch PM GrindControl User Manual

Page 33

Advertising
background image

Wireless Grinding Jar Cover

33

Fig. 13: Replacing the filter fleece

Replace the filter fleece.

Before assembly, check the state and the correct position of the seals and
the filter fleece.

Now, insert the cover baseplate back in the cover housing; when doing so,
pay attention to the position of the filter fleece (11).

Next, tighten the cover baseplate with the three screws (13).

Screw the antenna (4) onto the grinding jar cover. Tauschen Sie das
Filtervlies aus.

6.10 Replacing the Filter Fleece When Using the Gassing Function

Tools required:
13-mm open-jaw wrench
Hexagon socket screw key 2.5 mm:

Unscrew the antenna (4).

Remove the two de-aeration valves (2), using a 13-mm open-jaw wrench.

Remove the three screws (13) on the pre-assembled standard cover
baseplate and lift the plate up from the cover housing.

Replace the filter fleece (see fig. Replacing the filter fleece)

Before assembly, check the state and the correct position of the seals and
the filter fleece.

Now, insert the cover baseplate with the mounted de-aeration tubes in the
cover housing und when doing so, pay attention to the position of the filter
fleece (11).

Next, tighten the cover baseplate with the three screws (13).

Screw the two de-aeration valves (2) into the venting holes (9) and tighten
them with a 13-mm open-jaw wrench.

Screw the antenna (4) onto the grinding jar cover. Schrauben Sie die
Antenne (4) ab.

Pos : 8.20 /00020 BDA Softwar e/20005 PMGC Kapitels ammlung/Ü bersc hriften/1.1 Übersc hriften/11 M ahl becher sc hließ en @ 0 \mod_1214827689298_9.doc @ 1861 @ @ 1

6.11 Clamping the Grinding jar with the Clamping Ring

Pos : 8.21 /00020 BDA Softwar e/20005 PMGC Kapitels ammlung/War nhinweise/Hinweis nicht in PM 200 ver wendbar @ 0\mod_1210056716286_9.doc @ 766 @ @ 1

The wireless grinding jar cover cannot be used in the PM 200.

Advertising