Transport, scope of delivery, installation

17

•

Use the supplied connection cable to connect the device to the mains power
supply.

Pos : 8. 18 /0005 RET SCH/ 0005 R ETSCH Bedienungsanlei tungen Kapit elsamml ung en/AS200Jet/ 0006 AS200J et Transport, Li ef erumf ang und Auf st ellen/AS200J et M odul ext erne Absicherung @ 1 \ mod_1243926769314_9. doc x @ 10385 @ @ 1

The external fuse must be at least T15A (230V) T15A (100/120V).

Pos : 8. 19 /0025 War nhi nweis e/ W0015 WARNUNG Kabelbr uc h Stroms toß Er dung @ 4 \mod_1334064762171_9. doc x @ 27560 @ @ 1

WARNING

W0002

Danger to life through electric shock
–

An electric shock can lead to burns and to cardiac arrhythmias or to
respiratory arrest and cardiac arrest.

•

The device may only be operated with plugs that have a protective
conductor (earthed).