Model GX-86A Instruction Manual

RKI Instruments, Inc.

Page 23

Cylinder.

To change the calibration values, press the (SHIFT)/▼ and
DISPLAY/(ADJ.). The following screen will appear.

Press the AIR/▲ or (SHIFT)/▼ to adjust the LEL value to agree
with the concentration on the Four-Gas Cylinder. Press ON/OFF/
ENTER to advance to the SET CAL screen, the repeat the
adjustment procedure for O

2

, H

2

S and CO.

2. Perform Auto Calibration

After setting all Calibration Values, the display will return to the
Calibration Values screen.

a.Press ON/OFF/ENTER to go to the next screen:

b.Turn on the Four-Gas Cylinder and wait approximately 1 minute

for the flow and instrument readings to stabilize.

c.Press ON/OFF/ENTER to calibrate the detection circuits. The

AUTO CALIBRATION
50 12.0 25.0 50

SET CAL. <HC>

50 LEL

GAS/WAIT/ENTER

0 20.9 0.0 0