MAINTENANCE

Batteries

1. Check the battery voltage periodically by pressing the DISPLAY

switch to reach the Battery Voltage function. Replace the batteries
before the voltage drops to the operational limit (see Operation-
Alarms).

WARNING

T

AKE

THE

GX-86A

TO

A

NON

-

HAZARDOUS

LOCATION

BEFORE

CHANGING

OR

CHARGING

THE

BATTERIES

.

2. To replace the batteries, remove the battery compartment cover.

Remove the batteries and verify that the battery compartment and
electrical contacts are clean. Insert fresh batteries (alkaline or
fully-charged NiCd) according to the polarity (+/-) markings and
replace the cover.

3. Recharging Batteries

Rechargeable batteries may be recharged inside the GX-86A in a
non-hazardous location. Turn the instrument off, connect the
charger to the charger jack, and plug the charger into a 115 V AC
outlet. Full charge requires approximately 12 hours.

3. Bias Current Discharge

The batteries continuously supply a small current to maintain the
toxics sensors, even when the instrument is off (see Sensor
Maintenance). This current drain is minimal, but will result in a
normal discharge of the batteries over a period of several weeks.