Model GX-86A Instruction Manual

RKI Instruments, Inc.

Page 27

H

2

S and CO Span

1. Connect a H

2

S or CO calibration cylinder to the adapter.

2. From the calibration menu, scroll to “4.SET SPAN <H2S>,” or

“5.SET SPAN <CO>.” Press ON/OFF/ENTER to start the selected
calibration.

3. Adjust the calibration value following the procedure in Combusti-

bles Span, steps 2 - 6.

Note

The GX-86A will display up to 99 ppm H

2

S, but the limit of

linear response is 60 ppm, well above acceptable exposure
levels..

The GX-86A will display up to 499 ppm CO, but the limit
of linear response is 150 ppm, well above acceptable expo-
sure levels.

If any reading cannot be set to agree with the calibrating gas sample,
replace the sensor.