Model GX-86A Instruction Manual

Page 24

Hayward, California (510) 441-5656

END screen will appear:

d.After the end screen, the display automatically returns to the

first calibration screen.

e Turn off the Four-Gas Cylinder and disconnect it from the valve.

f. Press (SHIFT)/▼ to scroll through the options to “7.RETURN,”

then press ON/OFF/ENTER to return to normal operation.

Combustibles Span

Note

The combustible gas sensor is a general hydrocarbon sensor
that responds to most flammable vapors and gases; the
response will vary depending upon the substance. For best
results, calibrate the instrument to the gas or vapor intended
to be detected. See Appendix A for Relative Response
Curves for common gases and vapors.

1. At the Calibration Mode screen, press (SHIFT)/▼ to scroll to “2.

SET SPAN <HC>.” Press ON/OFF/ENTER to select. The display
will show:

2. Turn on the combustible gas calibration source, for example a

50% LEL cylinder. Wait approximately 1 minute while the LEL

AUTO CALIBRATION

NORMAL END

HC CAL. 0LEL

GAS/ADJ./ENTER