RKI Instruments GX-86A User Manual

Page 35

Advertising
background image

Model GX-86A Instruction Manual

Page 34

Hayward, California (510) 441-5656

USER OPTIONS

The advanced microprocessor program in the GX-86A allows the user
to select and adjust many of the detection and data logging features.
The GX-86A is pre-set to suit most applications; follow these
instructions only if required.

Entering Programming Mode

1. Take the GX-86A to a non-hazardous location—the instrument

does not detect gas during programming operations.

2. Turn the GX-86A off.

3. Hold down the AIR and (SHIFT) switches, then press ON. The

programming menu will appear:

General Programming Hints

• Press (SHIFT) to scroll down through the menu, and to change val-

ues in a specific option.

• Press AIR to scroll up through the menu, and to change values in a

specific option.

• Press ON/OFF/ENTER to select an individual option, and to enter

data during programming.

• The current option shows a flashing number.

• Many options have on-screen instructions

1.SET NVRAM DEFAULT
2.SET COMBINATION GAS