the start-up sequence and after the Demand Zero command.

3. The O

2

reading tends to drift with instrument orientation.

Toxics Sensors

Replace the sensor when:

1. The detection circuit cannot be calibrated correctly.

2. The display does not show 0 (CO) or 00 (H

2

S) immediately after

the start-up sequence and cannot be set to zero by Demand Zero.

Note

Allow up to 1/2 hour after the toxics sensors have been
replaced to show a normal response, then calibrate.

Sensor Replacement

1. Take the GX-86A to a non-hazardous location and turn the power

off.

2. Remove the instrument from the carrying case.

3. Press in the cover latch and slide the sensor cover down and away

from the instrument case.

4. To replace individual sensors, press down and turn counter-clock-

wise, then pull out of block. Figure 4. shows the installation of the