Chapter 3: installation & startup, Overview, Mounting the m2 transmitter – RKI Instruments M2 for 12 VDC User Manual

Page 13

Advertising
background image

M2 Transmitter Operator’s Manual, 12 VDC Operation

8

Chapter 3: Installation & Startup

Overview

This chapter describes procedures to mount the M2 Transmitter in the monitoring
environment and wire it to input power and devices.

Mounting the M2 Transmitter

1.

Select a mounting site that is representative of the monitoring environment. Consider
the following when you select the mounting site.

Select a site where the M2 is not likely to be bumped or disturbed. Make sure
there is sufficient room to perform start-up, maintenance, and calibration
procedures.

Select a site where the target gas is likely to be found first. For lighter gases,
mount the detector near the ceiling; for heavier gases, mount the detector near the
floor.

2.

At the monitoring site, use #10 screws through the junction box’s two mounting holes
to secure the junction box to a vertical surface. Figure 3 shows the outline and
mounting dimensions of the M2. Mount the M2 with the detector facing down as
shown in Figure 3.

Figure 3: Outline & Mounting Dimensions, LEL

3/4 NPT
Conduit Hub

1 1/2-20 For
Calibration Cup

7.7 max

2.3 max

5.23

5.00

2.41

.30 (2X)

.47 (2X)

4.59

5.86

1.02