M2 Transmitter Operator’s Manual, 12 VDC Operation

22

Chapter 7: Maintenance

Overview

This chapter describes procedures for performing preventive maintenance,
troubleshooting, calibrating the M2, and replacing field replaceable parts.

Preventive Maintenance

This section describes a recommended preventive maintenance schedule to ensure the
optimum performance of the M2. It includes daily, monthly, and quarterly procedures.

Daily

Verify a display reading of zero. Investigate significant changes in the reading.

Monthly

This procedure describes a test to verify that the M2 responds properly to the target gas.

NOTE:

Performing a response test on the M2 may cause alarms. If you do not want
external alarms to actuate, disable any external alarms during the response test
to avoid unwanted external alarm indications.

NOTE:

The following procedure assumes the use of a calibration kit which includes a
calibration gas cylinder, a fixed flow regulator with an on/off knob, a
calibration cup for the detector, and a short piece of sample tubing to connect
the regulator to the calibration cup.

1.

Disable all external alarms if external alarm indications are not desired during the
response test.

2.

Connect the calibration kit sample tubing to the calibration cup hose barb.

3.

Screw the calibration cup onto the detector.

4.

Connect the fixed flow regulator to the calibration cup using the sample tubing.

5.

Screw the calibration gas cylinder into the fixed flow regulator.

6.

Turn the regulator knob counterclockwise to open the regulator.

7.

Allow calibration gas to flow for one minute.

8.

Verify that the display reading is within ± 20% of the gas concentration.

NOTE:

If the reading is not within ± 20% of the gas concentration, calibrate the M2 as
described in “Calibration” on page 25.

9.

Turn the regulator knob clockwise to close the regulator.

10. Unscrew the calibration gas cylinder from the fixed flow regulator and unscrew the

calibration cup from the detector.