Appendix a: control button quick reference guide – RKI Instruments M2 for 12 VDC User Manual

Page 44

Advertising
background image

39

M2 Transmitter Operator’s Manual, 12 V DC Operation

Appendix A: Control Button Quick Reference Guide

The M2’s control buttons allow access to operational modes, resetting of alarms, and
display of the Information Screen. Table 14 shows which button combinations perform
these functions and which parameters are available for adjustment while in the
operational modes. While in these modes, display prompts showing a “?” require you to
respond by pressing either the UP/YES (for yes) or DOWN/NO (for no) button. To
change parameter settings, use the UP/YES and DOWN/NO buttons to get the desired
setting, then press and release the ENTER button to accept the setting.

Table 14: Control Button Quick Reference Chart

Mode Entered or

Function Performed

Button

Combination

Adjustable or Viewable Parameters

Alarm Reset

Press and
Release ENTER

n/a

Show Information
Screen

Press and Hold
DOWN/NO

• Input Voltage
• M2 Version/Revision Information

Calibration Mode

Press and Hold
UP/YES for 5
seconds

• Fresh Air Adjust
• Span (or Zero for Oxygen) Adjust with

Calibration Gas

Configuration Mode

Press and Hold
UP/YES &
ENTER for 5
seconds

• Alarm Levels
• Increasing or Decreasing Alarms
• Normally Energized or Normally De-

energized Relays

• Latching or Self-resetting Alarms
• Alarm On Delay
• Alarm Off Delay (if self-resetting)
• Zero Suppression
• Noise Filter
• Calibration Time-out Period

Gas Type Mode

Press and Hold
UP/YES and
DOWN/NO for
5 seconds

Gas Type

Modbus Mode

Press and Hold
DOWN/NO and
ENTER for 5
seconds

• Enable or Disable Modbus
• Slave ID
• Baud Rate
• Parity
• Response Delay