M2 Transmitter Operator’s Manual, 12 V DC Operation

Register 39

Register 39 is the auto-zero disable.
0=AutoZero Enabled
1=AutoZero Disabled

Register 40

Register 40 is the remote configuration register access level.
0=None
1=Alarm Reset

Function Code 16: Write Registers

There are 21 registers in Function Code 16 that can be used to write to the M2. This
manual only describes Register 16 because it can be used to reset an alarm condition. For a
complete description of Function Code 16, request Appendix C from RKI Instruments,
Inc.

Register 16

Register 16 performs the same function as the ENTER button on the M2. It is useful for
remotely silencing alarms. With the M2 in an alarm condition, bit [0] can be used to silence
it.

Table 12: Register 16, Button Function

NOTE:

Register 40 in Function Code 03, the remote configuration register, must be set
to 1 in order for Register 16 to work.

Bit

Value & Field

[0]

ENTER (also used for “Alarm Reset”)
0=Not Pressed
1=Pressed