Alarm indications – RKI Instruments M2 for 12 VDC User Manual

Page 19

Advertising
background image

M2 Transmitter Operator’s Manual, 12 VDC Operation

14

When the M2 is in its warm-up period, the signal output will be fixed at 3.5 mA (zero).

When the M2’s gas type is changed, the M2 will enter Configuration Mode for you to
verify the parameter settings. When you exit Configuration Mode, the display will
indicate NEEDS CALIBRATION and will continue to indicate this until Calibration
Mode is entered and a calibration is performed. In this situation, the signal output will
be fixed at 3.5 mA from the time Gas Type Mode is entered until the M2 is calibrated
and returns to normal operation.

If you enter Calibration Mode, Configuration Mode, Gas Type Mode, or Modbus
Mode, the signal output will be fixed at 3.5 mA until the M2 returns to normal
operation.

If the M2’s input power decreases below 10.5 volts so that the M2 is in a low power
alarm, the signal output is fixed below 2.4 mA until the low power alarm is cleared.

If the M2 goes into a fail condition, after a 30 second delay, the signal output is fixed
below 2.4 mA until the fail alarm is cleared. During the 30 second delay, the signal
output follows the detector output. In the case of a downscale reading, the display
and the signal output continue to track the reading down to -99% of full scale (1.15
mA).

Alarm Indications

NOTE:

The M2 includes alarm on and alarm off delay settings for alarm 1 and alarm 2.
The alarm indications described in this section operate according to the factory
set alarm settings. See Table 5 on page 18 for all the factory settings.

Table 4: Visual and Audible Alarm Indications

Condition

Cause

Visual Indication(s)

Alarm 1

1

Increasing gas reading at or above the
alarm 1 setpoint

• A1 LED is on
• Gas reading alternates

with ALARM-1 message

Alarm 2

1

Increasing gas reading at or above the
alarm 2 setpoint

• A2 LED is on
• Gas reading alternates

with ALARM-2 message

Fail

• Disconnected or misconnected

detector wiring

• Display reading at -10% of full

scale or lower

• Defective components

• F LED is on
FAIL message replaces

gas reading

NOTE: There is a 30 second
delay on the fail condition.

Low Power

DC power source less than 10.5 volts.

• F LED is on
LowPower message and

actual voltage of

incoming DC power

*

1

If the M2 is in both an alarm 1 and an alarm 2 condition, both alarm LEDs are on and the

display alternates between the gas reading and the ALMS 1&2 message.