RKI Instruments GD-K11D User Manual

Page 10

Advertising
background image

6 • Description

Model GD-K11D Operator’s Manual

Status lights

The display board includes five status lights. Table 3 lists each light and its function.

Program buttons

The display board includes five buttons. The MODE,

(PEAK),

, and SET (AL.P)

buttons are directly below the display screen. The TEST button is the red button near the
bottom right of the display board. Table 4 lists each button and its function.

Table 3: Model GD-K11D Status Lights

Light

Status

PW/TR
(power)

• live power is connected to GD-K11D and the

power switch is ON (solid light)

• fail condition (flashing light)

AL

(alarm)

• gas alarm condition (solid light)
• over range condition (solid light)

ZERO

• GD-K11D is prepared for adjustment of zero

setting (flashing light)

• GD-K11D accepted zero setting (solid light)

SPAN

• GD-K11D is prepared for adjustment of span

setting (flashing light)

• GD-K11D accepted span setting (solid light)

SKIP

GD-11D is in maintenance mode (flashing light)

Table 4: Model GD-K11D Program Buttons

Button

Function

MODE

Enters maintenance mode

(PEAK)

• Increases/changes the setting on the display
• Displays the peak reading

Decreases/changes the setting on the display

SET (AL.P)

• Prepares a setting to be changed
• Enters a new setting
• Displays the alarm point

TEST

Enters alarm test mode

Advertising