Model GD-K11D Operator’s Manual

Operation • 15

Adjusting the Alarm Type

You can program the Model GD-K11D’s alarm to activate on increasing readings or
decreasing readings. The factory-set alarm type is increasing.

1. Loosen the thumbscrew at the top of the housing.

2. Rotate the housing cover toward the bottom of the housing to gain access to the

interior of the Model GD-K11D.

3. Adjust the INT switch to the “7” position.

The current alarm type displays, and the SKIP light flashes. H (“high”) indicates an
increasing setting; L (“low”) indicates a decreasing setting.

4. Press the MODE button, and make sure 7 appears on the left side of the display

screen.

5. Use the

or

button to display the alternate alarm type.

6. Press the SET button to confirm the new alarm type.

The SPAN light turns on indicating that the Model GD-K11D accepted the new
setting.

7. Return the INT switch to the “0” position to return to normal operation.

The SPAN and SKIP lights turns of.

8. Rotate the housing cover up to its original position, then secure the cover to the

housing with the thumbscrew.