Model GD-K11D Operator’s Manual

Installation • 9

2. Use two screws through the slot cutouts at the top of the housing to secure the

housing to a vertical surface at the mounting site (see Figure 2.)

3. Use two screws through the mounting holes at the bottom of the housing to support

the installation of the Model GD-K11D (see Figure 2.)

Connecting the Sample Lines to the Model GD-K11D

1. Attach 4 x 6 mm PTFE teflon tubing to the gas in fitting. Place the opposite end of the

tubing at the sampling area.

CAUTION:

Avoid loops or slumps in the incoming sample line. To reduce response time, keep the

incoming sample line as short as possible.

2. Attach 4 x 6 mm PTFE teflon tubing to the gas out fitting. Route the opposite end of

the tubing to an open area where the sample can safely disperse.

Wiring the Model GD-K11D

This section describes procedures to connect AC power to the Model GD-K11D. This
section also describes

optional

procedures to connect a controller or recorder and external

alarm device to the Model GD-K11D. Figure 3 illustrates wiring connections to the
Model GD-K11D.

WARNING: Always verify that the Model GD-K11D’s power switch is in the OFF

position before you make wiring connections to the Model GD-K11D.

Figure 3: Wiring the Model GD-K11D

AC POWER

115V, 60 HZ.

GROUND

NEUTRAL

HOT

E

ALARM
DEVICE

CONTROLLER/

RECORDER

+

4-20

mA

output

POWER

SOURCE

CAUTION: If the alarm device draws more than 1 amp, use the
alarm contacts to control a relay with an appropriate contact rating.