Control Buttons • 17

The following types of data files are saved in the Data window:

• Calibration History Files

A calibration history file is saved for each instrument that has
been downloaded. It records the calibration information for
every calibration that was downloaded. The GX-2003 can
save information for up to 20 calibrations in its memory.

• Interval Trend Data Files

Interval trend data is logged at the interval time defined in the
GX-2003. Each logged point is an average reading over the
previous time interval.

• Alarm Trend Data Files

Alarm trend data is logged around an alarm event. The GX-
2003 can save up to 8 alarm trend data files in its memory.

• Alarm Event Files

Alarm event files record gas alarm events that have been
downloaded from instruments. The GX-2003 can save up to
100 alarm events in its memory.

• Trouble Event Files

Trouble event files record sensor failure and system failure
events that have been downloaded from instruments. The
GX-2003 can save up to 100 trouble events in its memory.

Advertising