30 • Downloading Data from the GX-2003

7. If the unit has been downloaded manually (Automatic

Download not selected) using the Complete Download or
Instrument Information commands, the unit’s instrument
information may be viewed after downloading by using the
Instrument Information button if the unit is still turned on
and connected to the computer. The instrument information
includes instrument parameters such as serial number and
alarm points, and also includes the results of the most recent
successful calibration for each channel. To view instrument
information for a unit, click the Instrument Information
button along the right side of the Download window. The
Instrument Information window displays. This screen cannot
be printed from the downloading software. If the unit is
turned off, the Instrument Information window will be blank.

Figure 19: Instrument Information Window