Calibrating A GX-2003 With the Data Logging Software • 58

6. Use the selection boxes under the Span column to select the

channels you want to calibrate.

NOTE: If installed, the TC % volume sensor cannot be

calibrated at the same time as the other sensors
because a multi-gas cylinder that includes % volume
methane is not available.

7. Click the Zero button. The software will perform a zero

adjustment setting the oxygen channel to 20.9% and all
other channels to 0.

Figure 39: Set Window with Calibration Frame

Set Auto Cal. values to match

Click boxes to select

channels for calibration

gas cylinder