Biolite multispectral light source, Lighting/darkroom/lens, Using the biolite functions – UVP Life Science User Manual

Page 95

Advertising
background image

Acquire Images

81

Lighting/Darkroom/Lens

BioLite MultiSpectral Light Source

The automated BioLite

TM

MultiSpectral Light Source is an added accessory to some systems. The

Automated BioLite has a fiber optic cable that threads into the darkroom and strictly controls the

wavelength of light shining inside the darkroom.

In the Acquisition Action Tab in the

Lighting menu is a list of functions for the BioLite. These are:

Filter: Drop down list consists of a variety of filters defined by the user.

Light Engine: Turn the BioLite On and Off.

Template: A template is a group of BioLite settings, which can be saved under a common name.

Select the Template for the sample, to apply from the drop-down box.

Intensity: Provides the range for light intensity. The intensity ranges from 0 to 100%.

Light Table: Turns the light on for the Transillumination light or for the Epi Illumination light.

Connect: Connects the Automated BioLite device from the software.

Using the BioLite Functions

For additional information on using the automated BioLite MultiSpectral Light Source, refer to the user

instruction manual for this unit.

Turn on the

Light Engine in the software.

Set the

Intensity of the BioLite. Select the highest intensity then move to a lower threshold if

desired.