Acquire Images

83

Darkroom Control

The software works with a series of darkrooms offered by UVP. The software controls the darkroom

lighting and other hardware inside the darkroom. If the system includes a motorized darkroom, the lift

platform height position will be controlled by the software as well. This software controls the following

systems:

BioSpectrum® Systems

iBox® Scientia Systems


Note: For iBox Explorer system settings, see separate information.

To access the darkroom controls, click onto the

Acquisition Action Tab > Lighting menu button.

Hardware Settings Selection

Filter: Click the drop down menu to select the user-defined emission filter.

Reset Wheel: Click on this button to realign the filter wheel after accidental movement of the

darkroom.

Epi Illumination: Darkroom supplies 365nm, 480nm and white light epi (overhead) illumination.

Click onto the radio button to make a selection.

Transillumination: Darkroom supplies power to the UV transillumination source and also to the

white light transillumination source. Click onto a radio button to make a selection.

Open Door Override: Under standard operating conditions, when the darkroom door opens, the

UV Light shuts off in the darkroom. When

UV Safety Off is selected as the door opens, the UV

Light stays on.

Tray Height

Open Door Override: Under normal circumstances when the darkroom door is open, the UV

light shuts down; to override this occurrence, click onto the Open Door Override switch.