Current sink mode, External voltage inputs, Current source mode (continued) – Weidmuller Portacal 1000 - Instrument Calibrator User Manual
Page 13
![background image](/manuals/575626/13/background.png)
7
i=2
On
Off
Off
Х Щ в б
Adjust Output Current to 2 mA
i=2
Ó
Accept output
i=20
Х Щ в б
Adjust Output Current to 20 mA
i=20
Ó
Accept output
Sure
Ó
To Save Values (or any other key to restart).
Current Sink Mode
Display
I+
I-
V
Keys
Action
Switch on Power, Calibrate link In CAL Position
u=0
Off
Off
On
Т Т
Press twice to select Current Sink mode
Connect Power Supply and Multimeter to measure output current (mA).
i=2
Off
On
Off
Х Щ в б
Adjust Output Current to 2 mA
i=2
Ó
Accept output
i=20
Х Щ в б
Adjust Output Current to 20 mA
i=20
Ó
Accept output
Sure
Ó
To Save Values (or any other key to restart).
External Voltage Inputs
Display
I+
I-
V
Keys
Action
Note: you must calibrate the Voltage Source Mode before completing this procedure.
Switch on Power, Calibrate link In CAL Position
u=0
Off
Off
On
Ф Х Щ Щ У
Press sequence to select External Voltage
mode. INC and DEC Leds will both light.
Connect voltage source and Set to 0 V.
U=0
Off
Off
On
Ó
Calibrate 0 V point
Current Source Mode (continued)