Current sink mode, External voltage inputs, Current source mode (continued) – Weidmuller Portacal 1000 - Instrument Calibrator User Manual

Page 13

Advertising
background image

7

i=2

On

Off

Off

Х Щ в б

Adjust Output Current to 2 mA

i=2

Ó

Accept output

i=20

Х Щ в б

Adjust Output Current to 20 mA

i=20

Ó

Accept output

Sure

Ó

To Save Values (or any other key to restart).

Current Sink Mode

Display

I+

I-

V

Keys

Action

Switch on Power, Calibrate link In CAL Position

u=0

Off

Off

On

Т Т

Press twice to select Current Sink mode

Connect Power Supply and Multimeter to measure output current (mA).

i=2

Off

On

Off

Х Щ в б

Adjust Output Current to 2 mA

i=2

Ó

Accept output

i=20

Х Щ в б

Adjust Output Current to 20 mA

i=20

Ó

Accept output

Sure

Ó

To Save Values (or any other key to restart).

External Voltage Inputs

Display

I+

I-

V

Keys

Action

Note: you must calibrate the Voltage Source Mode before completing this procedure.

Switch on Power, Calibrate link In CAL Position

u=0

Off

Off

On

Ф Х Щ Щ У

Press sequence to select External Voltage
mode. INC and DEC Leds will both light.

Connect voltage source and Set to 0 V.

U=0

Off

Off

On

Ó

Calibrate 0 V point

Current Source Mode (continued)