External current inputs, External voltage inputs (continued) – Weidmuller Portacal 1000 - Instrument Calibrator User Manual
Page 14
Advertising
8
Set voltage source to 10 V. Note: the error message
SHrt indicates a bad connection.
U=10
Off
Off
On
Ó
Calibrate 10 V point
Sure
Ó
To Save Values (or any other key to restart).
External Current Inputs
Display
I+
I-
V
Keys
Action
Switch on Power, Calibrate link In CAL Position.
u=0
Off
Off
On
Ф Х Щ У
Press sequence to select External Voltage
mode. INC and DEC Leds will both light.
Connect current source and Set to 0 mA.
I=0
On
Off
Off
Ó
Calibrate 0 mA point
Set current source to 20 mA. Note: the error message
Open indicates a bad connection.
I=20
On
Off
Off
Ó
Calibrate 20 mA point
Sure
Ó
To Save Values (or any other key to restart).
External Voltage Inputs (continued)