External current inputs, External voltage inputs (continued) – Weidmuller Portacal 1000 - Instrument Calibrator User Manual

Page 14

Advertising
background image

8

Set voltage source to 10 V. Note: the error message

SHrt indicates a bad connection.

U=10

Off

Off

On

Ó

Calibrate 10 V point

Sure

Ó

To Save Values (or any other key to restart).

External Current Inputs

Display

I+

I-

V

Keys

Action

Switch on Power, Calibrate link In CAL Position.

u=0

Off

Off

On

Ф Х Щ У

Press sequence to select External Voltage
mode. INC and DEC Leds will both light.

Connect current source and Set to 0 mA.

I=0

On

Off

Off

Ó

Calibrate 0 mA point

Set current source to 20 mA. Note: the error message

Open indicates a bad connection.

I=20

On

Off

Off

Ó

Calibrate 20 mA point

Sure

Ó

To Save Values (or any other key to restart).

External Voltage Inputs (continued)

Advertising