Yamato Scientific HC200 Cool Block User Manual
Page 39

37
Contents of a Segment
•
Other Elements relating to Segment
1. Wait Function (See the following example for setting the wait function).
If unexpected additional load or disturbance prevents measured temperatures from reaching the desired
temperature, this function keeps the operation in a block until the desired temperature is achieved. After
inputting the soak time, set ON or OFF.
2. End Segment
At the end of inputting a new program, set
in the ramp time of the last segment. When you have
composed a program up to the fifth segment, you do not need to set
Turning ON the wait function
3. Hold
If you set
in the soak time, the controller holds the desired temperature (ramp level) and closes the
program input mode.
If
is set, the holding time between 0 minute 0 second and 999 minutes flashes when you select the
soak time.
Indication when you have selected “Hold.”
! Be aware that this is different from the temperature holding function in the function menu.
Ramp Level
Push either the ▼ key
or ▲key.
999 minutes
0 minute 0 second