Warning, Requirements for installation – Yamato Scientific HC200 Cool Block User Manual

Page 7

Advertising
background image

5

REQUIREMENTS FOR INSTALLATION

Warning

Warning

Warning

Warning

Do not use the unit in an area where there is flammable or explosive gas.

Never use the unit in an area where there is flammable or explosive gas. The unit is not explosion-

proof. An arc may be generated when the power switch is turned ON or OFF, and fire/explosion may
result.

Always ground the unit.

Connect HC200’s power plug to a receptacle with grounding connectors.

Do not forget to ground HC200, to protect you and the unit from electrical shock in case of power surge.

Choose a receptacle with grounding connectors as often as possible.

Do not connect the grounding wire to a gas pipe, or by means of a lightning rod or telephone line. A fire

or electrical shock will occur.

If only bipolar receptacles are available for

HC200, connect an optional grounding
adapter to HC200’s power plug. Check
the polarity of the receptacle before
connecting the adapter to the receptacle.
Connect the adapter’s grounding wire
(green) to a grounding terminal to the power
supply. Contact our sales representative
in your vicinity or our service center for
additional information or assistance.

100V receptacle with grounding terminals

Power plug

100V bipolar receptacle

Grounding wire

Advertising