Collecting and Reviewing Data

Window

Type of window to be applied to the waveform measurement prior to
computing the spectrum

Hanning (default), Hamming, Flat Top, Rectangular

Averages

Sets the number of individual data sets to be incorporated into the
average calculation.

Numeric value
4 (default)

Overlap

The percentage of overlap during the averaging. It can be used to
reduce the total time that it takes to collect the data.

Percentage

Type

Sets the type of averaging for the measurement.

Spectral (default)
Time Synchronous
Peak Hold
Exponential

Ext Trig Slope

The input signal slope to be used for externally triggering the
measurement.
+ve - Defines the trigger event to occur on a rising edge.
-ve - Defines the trigger event to occur on a falling edge.

+ve (default)
-ve

Level

The amplitude level for externally triggering the measurement. The
level is a percentage of the trigger full scale range.

Numeric: 0.0 (default)

Bin Zeroing

Enables/disables bin zeroing. When set to Off, the bins below the
selected high pass filter are not zeroed. That is, data still appears but
is attenuated by the filter shape.
Set bin zeroing to On when you want to be sure unwanted
frequencies do not affect data. Set bin zeroing to Off when you want
to be aware of the presence of unwanted frequencies.

On (default)
Off

Table 43 - Offroute User Parameters

Parameter Name

Description

Values/Comments

Window

Description

Hanning

A general purpose window to use on random type data when frequency resolution is more important than
amplitude accuracy. Use this setting for most of your machinery monitoring activities.

Hamming

Another general purpose window with slightly different window characteristics than the Hanning
window.

Flat Top

Use this when amplitude accuracy is more important than frequency resolution. In data with closely
spaced peaks, a Flat Top window may smear the peaks together into one wide peak. Use this for
sinusoidal or calibration signals.

Rectangular

The Dynamix 2500 data collector does not apply a window. Use this only for transient signals that die out
before the end of the time sample, or for exactly periodic signals within the time sample.