Setting Up Measurements

3.

Finish setting up the dynamic proximity probe measurement by setting
Data Type to Magnitude and Units to mils or um.

You can use a Filter if desired.

Follow these steps to setup a Static (DC) component or gap.

1.

Set up a displacement transducer in the Transducer Calibration dialog box
with Input type set to DC Coupled and Units set to mils or um.

Make sure you enter the Calibration value that converts the voltage to a
displacement. If you are using the DC Input Module to remove the AC
component and attenuate the signal be a factor of 2.5, then you must
divide your transducer’s calibration by 2.5.

2.

Set up a Collection Specification with Transducer set to the DC
displacement and Signal detection set to RMS.

3.

Finish setting up the dynamic proximity probe measurement by setting
Data Type to Magnitude and Units to mils or um.

You must set the Filter to None because any type of high pass filtering
removes the DC component.

Typical Magnitude Measurement Definitions

This figure shows the setup for a typical magnitude measurement definition

Figure 15 - Magnitude measurement definition

The following table shows the STD (Hz) 2000 collection specification.

The following table shows the transducer specification in the STD (Hz) 2000
collection specification.

Table 24 - Collection specification for magnitude measurement definition

Transducer

Window

Signal Detection

Fmax

Lines

Phase

Order Norm?

Averages

Accelerometer

Hanning

Peak

2 kHz

400

No

No

4 linear

Table 25 - Transducer specification in STD (Hz) 2000 collection specification

Name

Base Unit

Input Type

Units

Calibration Value

DC Offset

Accelerometer

acceleration

ICP Accel

g’s

100

0