Chapter 2

Run Up/Coast Down Extension Module

Displaying Measurements
in Table Format

The magnitude and phase of selected orders at the varied rpm speeds are
displayed in a table.

Set Up the Table

1.

Select Table on the RUCD - Display Select screen and press F3 (Display).

The RUCD - Table Setup screen appears.

2.

Select a parameter press the Right arrow to open a list of values.

3.

Select or type in a parameter value.

4.

Press the Left arrow to save your selection.

Use the descriptions in

Table Parameter Descriptions on page 28

to help

you configure the parameters.

5.

When you are finished, press F4 (Display) to display the table.

See

Display the Table on page 30

.

Table 6 - Table Parameter Descriptions

Parameter Name

Description

Values/Comments

Start Spd (RPM)

Sets the start speed of the recorded data for the table.

This defaults to the Start Spd value set in the RUCD Setup screen.

See

Run Up/Coast Down Setup Parameters on page 18

.

10…3600 rpm

Stop Spd (RPM)

Sets the stop speed of the recorded data for the table.

This defaults to the Stop Spd value set in the RUCD Setup screen.

See

Run Up/Coast Down Setup Parameters on page 18

.

10…3600 rpm