Bayer HealthCare Ascensia Contour Blood Glucose Monitoring System User Manual

Page 25

Advertising
background image

21

What You See

What it Means

What You Should Do

HI

Test result is above

Wash your hands and test site. Repeat

600 mg/dL

the test. If the test result is still

(33.3 mmol/L)

“HI,” contact your physician or
healthcare professional (HCP)
immediately.
CAUTION: Glucose levels above
250 mg/dL (13.9 mmol/L) may
indicate a potentially serious medical
condition.

Control Test

• Control Solution is

Check all expiration dates. Do not

result is out-of-

past expiration date

use expired testing materials. Run

range (too high

or is past the 3-month another Control Test. If still out-of-

or too low).

open use date.

range, retest with a new Test Strip

• Test Strip is past

and Control Solution. If still out of

expiration date or is

range, call our Customer Service

past the 6 month

Department (1-800-348-8100).

open use date.

• Deteriorated Test

Strip due to heat,
cold or exposure to
moisture.

• Control Solution not

at room temperature.

Some display

Possible meter elec-

Replace batteries. If display is still

segments do

tronics or battery

incorrect, call our Customer

not appear

failure. Compare display

Service Department (1-800-348-8100)

during self-test.

with picture on inside

immediately. This may affect the

of Front Cover.

way you see your results.