Arterial blood evaluation (effect of po, Performance with arterial specimens) – Bayer HealthCare Ascensia Contour Blood Glucose Monitoring System User Manual

Page 32

Advertising
background image

1

Carl Burtis and Edward Ashwood (eds.): Tietz Textbook of Clinical Chemistry.
3rd Ed., W.B. Saunders Co. Philadelphia, PA. 1999, p. 1827.

Arterial Blood Evaluation (Effect of pO

2

:

Performance with Arterial Specimens)

An arterial specimen study was performed at a hospital site.
Ascensia C

ONTOUR

results from arterial samples were

compared to the site blood glucose method. The glucose
concentration of the samples ranged from 55 to 432 mg/dL,
with an average of 153 mg/dL. The sample hematocrit range
was 22 to 58%, with an average of 39%. The pO

2

values

ranged from 43 to 205 mm Hg, with an average of 104 mm
Hg. The reference range for arterial pO

2

is 83 to 108 mm Hg

(11.1 to 14.4 kPa)

1

. Mean biases were –2.9% at pO

2

concentrations <108 mm Hg (n = 33, Mean pO

2

= 71 mm Hg),

–2.5% at pO

2

concentrations between 83 to 108 mm Hg

(n = 40, Mean pO

2

= 94 mm Hg), and –0.8% at pO

2

concentrations >200 mm Hg (n = 38, Mean pO

2

= 143 mm

Hg). Two samples had glucose concentrations <75 mg/dL and
the bias was calculated as an absolute difference rather than
a percentage. These samples had biases of –1.0 mg/dL and
–8.0 mg/dL; pO

2

values were 80 and 112 mm Hg, respectively.

28

Advertising