Multiple Site Testing Evaluation

Forty persons with diabetes, in steady state for glucose,
performed self-capillary blood glucose assays using the
Ascensia C

ONTOUR

System and one reagent lot. A fingerstick

was performed with the Microlet Adjustable Lancing Device.
Samples from the palm, forearm, thigh, and abdomen were
obtained using the Ascensia™ M

ICROLET

V

ACULANCE

®

Lancing

Device. Ascensia C

ONTOUR

results from the multiple sites were

compared to those obtained from the finger. One subject was
unable to obtain a sample from the abdomen. The blood
glucose range from the fingerstick measurements was 49 to
407 mg/dL, with an average of 166 mg/dL. The hematocrit
range was 36 to 52%, with an average of 42%.