Accuracy, Bayer accuracy evaluation, High control solution – Bayer HealthCare Ascensia Contour Blood Glucose Monitoring System User Manual

Page 28: Normal control solution

Advertising
background image

24

Accuracy

1. Bayer Accuracy Evaluation:

To assess the accuracy of the Ascensia C

ONTOUR

System, 60

fresh fingerstick specimens from people with diabetes were
tested with the Ascensia C

ONTOUR

System (12 – 14 readings

per fingerstick) and the YSI 2300 STAT Plus Glucose Analyzer.
The glucose concentration of the samples ranged from 45 to
454 mg/dL with an average of 150 mg/dL. The range of
hematocrits was 32% to 49% with an average of 40%.

High Control Solution

Duplicate

Overall Overall

Reagent

Assays

Mean

SD

CV

Lot

Operator

(n = )

(mg/dL)

(mg/dL)

(%)

22B

HCP

(n = 1)

35

337.2

13.28

3.9

Lay User

(n = 51)

76

341.5

14.94

4.4

22C

HCP

(n = 4)

39

351.9

8.98

2.6

Lay User

(n = 53)

78

351.2

11.16

3.2

23B

HCP

(n = 5)

74

345.6

10.41

3.0

Lay User

(n = 104)

154

324.8

9.44

2.8

Normal Control Solution

Duplicate

Overall Overall

Reagent

Assays

Mean

SD

CV

Lot

Operator

(n = )

(mg/dL)

(mg/dL)

(%)

22B

HCP

(n = 1)

35

119.6

4.58

3.8

Lay User

(n = 51)

76

120.6

3.90

3.2

22C

HCP

(n = 4)

39

122.0

3.23

2.6

Lay User

(n = 53)

78

122.3

4.88

4.0

23B

HCP

(n = 5)

74

120.9

3.04

2.5

Lay User

(n = 104)

154

120.1

3.67

3.1