Bayer HealthCare Ascensia Contour Blood Glucose Monitoring System User Manual

Page 37

Advertising
background image

33

Reagent

95% Confidence Interval

Lot

Opr.

n

y =

Slope

/

Intercept

r

22B

HCP

56

0.90x + 2.78

0.85 to 0.97 –4.59

to 5.91 0.984

22C

HCP

56

0.90x + 2.39

0.86 to 0.96 –3.41

to 7.43

0.988

23B

HCP

56

0.89x + 3.84

0.85 to 0.94 –1.70

to 8.73

0.983

Opr. = Operator; HCP = Healthcare Professional

Passing and Bablok Regression Statistics / Pearson correlation