BOWERS METROLOGY page 14 /32 version 2.06 -E
3.4 PROGRAMMING A DIAMETER
Measurement of a diameter with a HBT probe connected to channel A.
Step n°1 is to place the probe into a location to be able to get the maximum range and the minimum
deviation.
Place the measuring head of 15.0000 mm in the master,
Press simultaneously
Move the probe down to the electro-mechanical zero. , then approximately stop around
zero, lock the probe . Press to escape to the measuring mode.
Parameters to be changed for example number 2
CHANNEL : Direc. -
MEASURE : M1 –Range :
± 0.005 mm Resol.: 0.0001mm –
SPEC. : M1 – NOMI. 015.0000 - -TOLER –0.0020 - +TOLER +0.0020 –APPROCH –0.0016
+ APPROCH : +0.0016
MASTER : M1 - -Master : 14.9988mm - +Master : 15.0026mm
Component
15.000 mm ± 0.002
Min master
14.9988mm
Max. master.
15.0026mm
S
Enter
Enter
Esc
A D J U S T
-
0 . 0 0 1 0
C H - A
- 1 . 0 5 0 0