Pc-2200 – Bowers Group Advanced Electronic/Air Column PC-2200 User Manual
Page 16

PC-2200
BOWERS METROLOGY page 16 /32 version 2.06 -E
The Calibration
Example N°2
The parameters are now stored. The automatic calibration can be started.
place the measuring head in the lower master.
place the measuring head in the upper master.
The sensitivity is calculated automatically.
If the message appears, that means that the value between the two masters is bigger
then the calculated value.
NONE : All sensitivity values are allowed.
000% : The sensitivity value will not be accepted if the calculated value during the calibration is different
from the stored value.
XXX% : The sensitivity value will not be accepted if the calculated value during the calibration is different
from XXX% of the stored value.
After the calibration the measurement can be started.