5.13 Manual Control

■■■■■■■■■■■■■■■■■■■■■■

48

■■■■■■■■

877 Titrino plus

5.13.2

Measuring

Open the dialog for manual measurement as follows:

1 Open the manual control

In the main dialog, select Menu and press [OK].
The main menu opens.

Select the menu item Manual control and press [OK].

Manual control opens.

2 Select a measuring mode

Select the Measure item.

In the function bar, select the measuring mode and press [OK].

3 Configure the measuring mode

Select the desired electrode from the sensor list.
The selection depends on the measuring mode. Sensors are
defined under System

Sensors.

Enter the measuring temperature if no temperature sensor is con-
nected. If a temperature sensor is connected then the tempera-
ture will be measured automatically.
This temperature is used for automatic temperature compensation
with pH measurements.

4 Start the measurement

Press [START].