Metrohm 877 Titrino plus User Manual

Page 85

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

7 Parameters

877 Titrino plus

■■■■■■■■

75

Measuring mode Upol:
Input range

–125.00 - 125.00 µA

Default value

–125.00 µA

Upper limit

This parameter is visible only when Window = on.

Measured value for the upper limit.

Measuring mode pH:
Input range

–20.000 - 20.000

Default value

20.000

Measuring mode U, Ipol:
Input range

–1250.0 - 1250.0 mV

Default value

1250.0 mV

Measuring mode Upol:
Input range

–125.00 - 125.00 µA

Default value

125.00 µA

EP criterion

The equivalence point criterion found (ERC = Equivalence point Recogni-
tion Criterion) is compared with this value. Equivalence points whose ERC
is less than the value defined here will be ignored.

Measuring mode pH:
Input range

0.10 - 9.99

Default value

0.50

Measuring mode U, Ipol:
Input range

1 - 999 mV

Default value

30 mV

Measuring mode Upol:
Input range

0.1 - 99.9 µA

Default value

2.0 µA

EP recognition

This parameter allows you to filter out only the equivalence points that are
being sought.

for Window = off
Selection

all | greatest | last | off

Default value

all

Advertising