3 ph adjustment, Ph adjustment – Metrohm 845 Eluent Synthesizer User Manual

Page 27

Advertising
background image

3.3 Method

845 Eluent Synthesizer / Instructions for Use 8.845. 8001EN

21

Total Volume mL [ 200...5000 ; 200 ]

Volume (in mL) to be prepared in the course of a mixing
process.

Changing the

Total Volume does not affect the ratio

of the

Components mixed. The numerical values of

the

Components given with the (absolute) unit mL are

adapted automatically.

Stirrer Speed [ -15...15 ; 4 ]

Speed of the stirrer.

/

Switch on / switch off stirrer.

When starting the mixing process (by clicking on

)

the stirrer is always switched on.

Opens the window

PH-ADJUSTMENT CONFIGURATION

.

3.3.3 pH

Adjustment

General information about pH adjustment

pH adjustment is always carried out after the main component has
been added (i.e. at the end of the mixing process).

For automatic pH adjustment

702 Titrinos from Metrohm are intended

for use. If the pH needs to be adjusted from both sides (without

changing the

pH-Controller liquid), two 702 Titrinos are required as

pH-Controllers.

In the fields

pH-Adjustment

and

pH-Controller

define whether and how

the pH is to be adjusted.

If

autonomous is selected for pH-Adjustment then none must not

be selected for

pH-Controller. For autonomous a pH-Controller

must be used.

Window "pH-Adjustment Configuration"

The dialog window

PH-ADJUSTMENT CONFIGURATION

can be opened

by clicking on the button

in the Main Window.

In the

PH-ADJUSTMENT CONFIGURATION

window the pH adjustment

procedure can be defined.