Figure 13, Installing tubing weighting and aspiration filter, Figure 14 – Metrohm 850 Professional IC Anion MCS User Manual

Page 38: Eluent aspiration tubing fully equipped

Advertising
background image

2.8 Eluent

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

850 Professional IC – Anion – MCS

4 Installing tubing weighting and aspiration filter

1

2

3

4

5

Figure 13

Installing tubing weighting and aspiration filter

1

Eluent aspiration tubing 6.1834.080

2

Eluent bottle attachment 6.1602.160

3

Tubing weighting
From accessory set 6.2744.210.

4

Clamping screw
From accessory set 6.2744.210.

5

Aspiration filter 6.2821.090
With filter holder from accessory set
6.2744.210.

Slide the tubing weighting (13-3) onto the eluent aspiration tub-
ing (13-1).

Slide the clamping screw (13-4) onto the eluent aspiration tubing
(13-1).

Insert eluent aspiration tubing (13-1) into the aspiration filter
(13-5). The end of the tubing must touch the base of the filter.

Screw together clamping screw (13-4) and filter holder (12-1).
The end of the tubing must still touch the base of the filter.

Figure 14

Eluent aspiration tubing fully equipped.

5 Mounting eluent aspiration tubing to the eluent bottle

Insert the eluent aspiration tubing into the eluent bottle (15-10).

Advertising