Figure 26, Msm – connections – Metrohm 850 Professional IC Anion MCS User Manual
Page 57

■■■■■■■■■■■■■■■■■■■■■■
2 Installation
850 Professional IC – Anion – MCS
■■■■■■■■
47
1
3
2
3
4
5
6
7
8
2
1
Figure 26
MSM – connections
1
Union nut
2
MSM connecting piece 6.2832.010
3
Eluent input capillary
Labeled with Eluent.
4
Eluent output capillary
Labeled with Detector.
5
Rinsing solution input capillary
Labeled with H2O.
6
Rinsing solution output capillary
Labeled with Waste.
7
Regeneration solution output capillary
Labeled with Waste.
8
Regeneration solution input capillary
Labeled with H2SO4.
The PTFE capillaries firmly mounted on the MSM are connected to the
other components of the IC system as follows:
Caution
As the PTFE capillaries are very soft, the pressure screws should not be
overtightened.
Flattened capillaries must be shortened with the help of a 6.2621.080
capillary cutter.