Figure 15, Eluent bottle – connected – Metrohm 850 Professional IC Anion MCS User Manual
Page 39

■■■■■■■■■■■■■■■■■■■■■■
2 Installation
850 Professional IC – Anion – MCS
■■■■■■■■
29
■
Fasten the bottle attachment (13-2) on the eluent bottle (15-10).
The aspiration filter (15-6) must rest on the base of the eluent
bottle.
6 Mounting the adsorber tube
Note
In the case of alkaline eluents and eluents with lower buffer
capacity, the eluent bottle must be equipped with a CO
2
adsorber
■
First, place a piece of wadding (15-3), then the CO
2
adsorber
(15-4) in the large opening of the adsorber tube (15-2) and close
with the plastic cover.
■
Fasten the adsorber tube (15-2) using the SGJ clip (15-12) onto
the bottle attachment (15-11).
1
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Figure 15
Eluent bottle – connected
1
Eluent aspiration tubing 6.1834.080
For aspirating the eluent. Pre-installed.
2
Adsorber tube 6.1609.000
3
Wadding
4
CO
2
adsorber
Adsorbs CO
2
from the air (e.g. Merck soda
lime with indicator, no. 6839.10).