6.7 871 Advanced Bioscan
IC Net 2.3
107
Startup values
E
Electric potential applied to the amperomet-
ric detector cell in V
Range:
-2 ... +2 V
.
Range
Selection of measuring range:
pA
10 ... 500 pA
.
nA
1 ... 500 nA
.
µA
1 ... 200 µA
.
Filter
Electronic damping of the analog output
signal.
Selection:
off, 0.2, 0.5, 1, 2, 5 s
Offset
Offset of the zero point for the analog out-
put signal.
Selection:
-50 ... +50 %
(in steps of 10 %)
Data rate
Setting of the data acquisition rate.
Selection:
-1, 2, 5, 10 Hz
Polarity
Set sign:
+
or
-
Actual
E
Display of current cell potential in V
Range
Display of current measuring range.
Filter
Display of current damping constant.
Offset
Display of current offset.
Data rate
Display of current data rate.
Polarity
Display of current polarity.
<Set>
Send current parameters immediately to the
871 Advanced Bioscan. Parameters are not
stored in the system file (
*.smt
) as long as
the file is not saved.
Pulse mode
This subtab of the
Method parameters
tab in the
871 Advanced Bio-
scan
window contains the method parameters for the Pulse
mode. These values are sent and applied to the 871 Advanced
Bioscan by
Start determination
and
Startup hardware
, or if the val-
ues are sent manually with
<Set>
.
Startup values
E1, E2, E3
Definition of three different working poten-
tials, which are applied to the amperometric
detector cell cyclically. The set potentials
are graphically displayed if the
Method
op-
tion is enabled above the graphic window.
Range:
-2 ... +2 V
.
t1, t2, t3
Time duration to apply the working poten-
tial.
Range:
0.1 ... 2 s
(t1)
, 0.01 … 2 s
(t2,t3)