5 spectrum window – Metrohm IC Net 2.3 User Manual
Page 272

6 Devices
IC Net 2.3
262
Operating hours
Operating hours of the lamps.
UV lamp, h
Display of the operating hours of the UV
lamp.
<Reset UV lamp>
Set
operating hours of the UV lamp
to 0. Click
it after replacing the UV lamp.
VIS lamp, h
Display of the operating hours of the VIS
lamp.
<Reset VIS lamp>
Set
operating hours of the VIS lamp
to 0. Click
it after replacing the VIS lamp.
6.28.5 Spectrum
window
The
SPECTRUM WINDOW
can be opened with a right-hand mouse-click on
the 844 icon and then selecting the menu item
Spectrum window
, or via the
button on the subtab
Manual/Detector
.
A selection can be made from three different spectra, the absorption
spectrum, intensity spectrum and the zero spectrum. Please note that each
time you click on
Absorption
,
Intensity
or
Zero
the function of the
<Update>
button will be triggered.
<Update>
The 844 measures the current intensity spectrum and
sends it together with the last zero spectrum to have
been recorded to <IC Net>. The absorption spec-
trum is recalculated from these two spectra.
Absorption
The absorption (with the units mAU) is given on the y-
axis as a function of the wavelength (x-axis). The ab-
sorption values are calculated for all measured wave-
lengths and shown a s spectrum. The calculation for-
mula for the absorption value is shown here as an ex-